Applications
Data logger with integral USB connector. Ideal for cost-effective monitoring of temperature sensitive goods in both short haul and long-distance transportation applications
Product Summary
The LogTag® USRIC-8 PDF Temperature Recorder measures and stores up to 8,000 real time temperature readings over a measurement range of -30 °C to +60 °C
Designed as a single-use, cost effective & reliable temperature recorder, the LogTag® USRIC-8, enclosed in a robust & durable polycarbonate case, automatically generates a fully detailed PDF report and provides the option to download the data using the freely available LogTag Analyzer software.
Using LogTag® Analyzer, the LogTag® USRIC-8 is easily set-up for a number of recording conditions.
Product Highlights
- Integral USB connector - No separate Interface Cradle required
- Automatically generates a PDF report - No special software required
- Encrypted LTD file generated for a more detailed analysis
- User configurable - No need to buy a special logger for each trip
- Suitable for independent receivers, clinical trials & last-mile distribution
Technical Specifications
Model | USRIC-8 |
Measurement Range | -30 °C ~ +60 °C |
Resolution | 0.1 °C |
Accuracy | Better than ±0.5 °C for measurements from - 5 °C~+30 °C Better than ±0.8 °C for other measurements |
Sensor Reaction Time | Typically less than 5 minutes (T90) in moving air (1 m/s) |
Capacity | 8,001 real time temperature recordings, non-volatile memory |
Sampling Frequency | Configurable from 1 minute to 18 hours |
Logging Start Options | Push button start or specific date/time |
Recording Indication | Flashing 'OK' indicator / flashing 'ALERT' indicator. |
Download Time | Typically with full memory (8000 readings) in less than 20 seconds from time of insertion to availability of PDF report. Typically less than 10 seconds from time of insertion to availability of LTD file in LogTag® Analyzer (if configured) |
Environmental | IEC 60529: IP64 with USB cap fitted |
Power Source | 3 V Lithium battery– non replaceable |
Alarm Functions | High/low, single events & cumulative supported, ‘OK’/’ALERT’ LEDs |
Connection Interface | USB 2.0, A-type plug |
Software | Adobe Reader, LogTag Analyzer® 2.5 or higher |
Battery Life | Projected life of 6 months of operation from time of configuration. Shelf life before configuration of up to 2 years. |
Size | 93 mm (H) x 54.5 mm (W) x 8.6mm (T) including protective USB cap. |
Weight | 32 g |
Case Material | Polycarbonate |
LogTag® Recorders reserves the right to change this product’s specifications without notice. |

EMC Compliance
Tested & complies with EC EMC directives (EN 50081-1:1992 & EN 61000-6-1:2001).
Tested & complies FCC Part 15 Subparts A & B.

FDA CFR21 Part 11 Compliance
Designed for FDA CFR21 Part 11 Digital signatures.

RoHS Directive
The RoHS Directive stands for "the restriction of the use of certain hazardous substances in electrical and electronic equipment". This Directive bans the placing on the EU market of new electrical and electronic equipment containing more than agreed levels of lead, cadmium, mercury, hexavalent chromium, polybrominated biphenyl (PBB) and polybrominated diphenyl ether (PBDE) flame retardants.

- Data Loggers
- LogTag® TRIX-8 / TRIX-16
- LogTag® TREX-8
- LogTag® HAXO-8LogTag® HASO-8LogTag® SRIC-4
- LogTag® UTRIX-16
- LogTag® USRIC-8
- LogTag® USRIC-4
- LogTag® USRID-16
- LogTag® UTRID-16
- LogTag® UHADO-16
- LogTag® UTRED30-16
- LogTag® UTRED30-WiFi
- LogTag® TRIL-8 / SRIL-8
- LogTag® TREL-8
- LogTag® TREL30-16
- LogTag® TRID30-7
- LogTag® TRED30-16R
- LogTag® vaxtag