Kontakt

Questions? Price info? Technical details?

Contacto

LogTag® TRED30-16CP Temperature Monitoring at a Glance

LogTag® TRED30-16CP Temperature Monitoring at a Glance

Applications

Temperature measurement where access requires a remote probe and data needs to be accessed without a PC.

 

Product Summary

The LogTag® TRED30-16CP is the Smart Probe ready iteration of the LogTag® TRED30-16R. When paired with the compatible LogTag® CP110 Smart Probe, you remove the need to relocate or return the LogTag® TRED30-16CP temperature recorder for calibration, ensuring continuous and uninterrupted logging.
The LogTag® TRED30-16 Temperature Logger measures and stores up to 15,900 temperature readings over a -40°C to +99°C (-40°F to +210°F) measurement range from either a smart Probe.

The display is designed to show ‘at a glance’ if temperature excursions have occurred during the current day and up to the previous 29 days. The display also shows the current temperature reading, the current time, recording status and battery status.


The alarm is triggered if temperature readings are outside pre-set alarm limits. The status indicator changes from a ✔ to a X and, if configured, the audible alarm will sound. The day marker will also appear if not already present. The alarm status indicator and audible alarm can be easily cleared, but the day marker remains to ensure a visible record of excursions is shown on the display. A simplified alarm clearance, stop & start procedure is available as an option.


Details of any excursion in the preceding 30 days can be checked directly by inspecting the statistics history on the recorder’s display or in more detail by downloading the logged data via a standard LogTag® Interface cradle to LogTag® Analyzer operating software.

Please note: An interface is required for configuration and reporting

 

Product Highlights

  • Easy to read LCD display with 30 day summary.
  • Built-in audible alarm.
  • Review button allows review of current day and up to 29 previous days of statistical data and also adds an ‘inspection mark’ into the recordings.
  • Flexible start options push button recording start with optional delayed start (up to 18 hours) or specific time & date start.
  • User replaceable coin cell battery.
  • 15,900 recordings capacity enough memory for 66 days @ 6 min logging.
  • Calibratable external temperature sensor (Smart Probe or ST100 sensor)

External Probes


Technical Specifications

Product model TRED30-16CP (replaceable battery)
External temperature sensor measurement range -40 °C ~ +99 °C
Operating temperature range  -30 °C ~ +60 °C
Storage temperature range -10 °C ~ +40 °C
Rated temperature reading accuracy

Better than ±0.5 °C for -20 °C~+40 °C - typically ±0,3 °C
Better than ±0.8 °C for -40 °C~+20 °C - typically ±0,5 °C
Better than ±0.8 °C for +40 °C~+70 °C - typically ±0.7 °C
Better than ±1.2 °C for +70 °C~+99 °C - typically ±1 °C

Rated temperature reading resolution on LCD 0.1 °C for measurements -40 °C ~ +50 °C
0.2 °C(0.4 °F) for measurements +50 °C ~ +70 °C
0.3 °C(0.5 °F) for measurements +70 °C ~ +80 °C
0.4 °C-0.6 °C for measurements +80 °C~ +99 °C
Recording capacity Data logging memory : 15,905 logs - eg 66 days @ 6min logging, 80 days @ 15min logging
Day summary statistics memory (for display on LCD): up for 30 days of Max/Min/duration and alarm trigger statistics
Sampling Interval configurable from 30 seconds to hourly
Logging start options Supports ‘continuous’ wrap around or specific recording period
Recording indication State indicator "REC"
Download time Typically less than 10 seconds for full memory (15,905 readings), depending on computer or readout device used.
Environmental IP61 (when vertically mounted or hung)
Power source 3 V CR2032 (LiMnO2) coin cell battery, user replaceable
Battery life Typically 1 year of operation.
NOTE: Typical battery life assumes 6 minute logging, day statistics are reviewed on the display no more than once daily for no longer than 30 seconds each time.
Download data monthly.
NOTE: Continual activation of the audible alarm will reduce the working life of the battery. When an alarm is triggered, the alarm should be cleared and reset as
soon as possible.
Real time clock Built-in real time clock.
Rated accuracy ±25 ppm @ 25 °C (equivalent to 2.5 seconds/day).
Rated temperature coefficient is -0.034 ±0.006ppm/°C (I.e typically +/- 0.00294 seconds/day/°C).
External probe cable lengths Standard: 1.5 m.
Extended: 3 m (recommended maximum).
Remote external probe type PTFE (FDA food contact rated) coaxial.
Size 93.0 mm(H)x54.5mm(W)x8.6mm(T)
Weight 43 grams without external probe
Case Material Polycarbonate
EMC Compliance Complies with EC EMC directives (EN 61000-6-3:2006 & EN 61000-6-1:2005)
Complies with FCC Part 15 Subparts A and B.
FDA 21CFR Part 11 Compliance Designed to comply with FDA 21CFR Part 11 (Digital signatures)
Reader interface  Standard LogTag® Interface cradle
Manufactured to meet the European RoHS Directive
LogTag® Recorders reserves the right to change this product’s specifications without notice.