Applications
Temperature measurement where access requires a remote probe and data needs to be accessed without a PC.
Product Summary
The LogTag® TRED30-16R Temperature Recorder measures and stores up to 15,900 temperature readings over a -40 °C to +99 °C measurement range, using a remote temperature probe.
The display is designed to show ‘at a glance’ if temperature excursions have occurred during the current day and up to the previous 29 days. The display also shows the current temperature reading, the current time, recording status and battery status.
The alarm is triggered if temperature readings are outside pre-set alarm limits. The status indicator changes from a ✔ to a X and, if configured, the audible alarm will sound. The day marker will also appear if not already present. The alarm status indicator and audible alarm can be easily cleared, but the day marker remains to ensure a visible record of excursions is shown on the display. A simplified alarm clearance, stop & start procedure is available as an option.
Details of any excursion in the preceding 30 days can be checked directly by inspecting the statistics history on the recorder’s display or in more detail by downloading the logged data via a standard LogTag® Interface cradle to LogTag® Analyzer operating software.
Please note: An interface is required for configuration and reporting
Product Highlights
- Data logging Records time and temperature data for later download.
- State Indicators Display shows current state (READY,STARTING,RECORDING, PAUSED or STOPPED).
- Built-in Audible Alarm
- Alarm Symbol Appears on the display to show an alarm trigger has occurred.
- Day alarm summary Displays in calendar-like format alarm triggers for up to the last 30 days.
- Current time & temperature Shows the current time and the last recorded temperature when not in review mode.
- Review button Allows review of current day and up to 29 previous days of statistical data and also adds an ‘inspection mark’ into the recordings.
- Start/Clear/Stop button Push button recording start, alarm clear and recording stop (each option separately configurable).
- Flexible start options Push button recording start with optional delayed start (up to 18 hours) or specific time & date start.
- Flexible recording options Either record for a fixed time or record continuously with ‘wrap around logging’.
- Rapid Download! Takes only seconds to download recordings.
- Thin flat case Fits easily into packaging and thin enough to be easily mailed “letter rate”.
- Real time clock Records time & temperature simultaneously.
- 15.900 recordings capacity Enough memory for 66 days @ 6 min logging.
- High Quality Connector Gold plated remote probe connector.
- Interchangeable Probes LogTag® remote temperature ST100 probes are interchangeable.
Technical Specifications
Product model | TRED30-16R (replaceable battery) |
External temperature sensor measurement range | -40 °C ~ +99 °C |
Operating temperature range | -30 °C ~ +60 °C |
Storage temperature range | -10 °C ~ +40 °C |
Rated temperature reading accuracy |
Better than ±0.5 °C for -20 °C~+40 °C - typically ±0,3 °C |
Rated temperature reading resolution on LCD | 0.1 °C for measurements -40 °C ~ +50 °C 0.2 °C(0.4 °F) for measurements +50 °C ~ +70 °C 0.3 °C(0.5 °F) for measurements +70 °C ~ +80 °C 0.4 °C-0.6 °C for measurements +80 °C~ +99 °C |
Recording capacity | Data logging memory : 15,905 logs - eg 66 days @ 6min logging, 80 days @ 15 min logging Day summary statistics memory (for display on LCD): up for 30 days of Max/Min/duration and alarm trigger statistics |
Sampling Interval | configurable from 30 seconds to hourly |
Logging start options | Supports ‘continuous’ wrap around or specific recording period |
Recording indication | State indicator "REC" |
Download time | Typically less than 10 seconds for full memory (15,905 readings), depending on computer or readout device used. |
Environmental | IP61 (when vertically mounted or hung) |
Power source | 3V CR2032 (LiMnO2) coin cell battery, user replaceable |
Battery life | Typically 1 year of operation. NOTE: Typical battery life assumes 6 minute logging, day statistics are reviewed on the display no more than once daily for no longer than 30 seconds each time. Download data monthly. NOTE: Continual activation of the audible alarm will reduce the working life of the battery. When an alarm is triggered, the alarm should be cleared and reset as soon as possible. |
Real time clock | Built-in real time clock. Rated accuracy ±25 ppm @ 25 °C (equivalent to 2.5 seconds/day). Rated temperature coefficient is -0.034 ±0.006 ppm/°C (I.e typically +/- 0.00294 seconds/day/°C). |
External probe cable lengths | Standard: 1.5 m. Extended: 3 m (recommended maximum). |
Remote external probe type | PTFE (FDA food contact rated) coaxial. |
Size | 93.0 mm(H)x54.5 mm(W)x8.6 mm(T) |
Weight | 43 grams without external probe |
Case Material | Polycarbonate |
EMC Compliance | Complies with EC EMC directives (EN 61000-6-3:2006 & EN 61000-6-1:2005) Complies with FCC Part 15 Subparts A and B. |
FDA 21CFR Part 11 Compliance | Designed to comply with FDA 21CFR Part 11 (Digital signatures) |
Reader interface | Standard LogTag® Interface cradle |
![]() |
Manufactured to meet the European RoHS Directive |
LogTag® Recorders reserves the right to change this product’s specifications without notice. |
NEW!

- Data Loggers
- LogTag® TRIX-8 / TRIX-16
- LogTag® TREX-8
- LogTag® HAXO-8LogTag® HASO-8
- LogTag® SRIC-4
- LogTag® UTRIX-16
- LogTag® EnviroLog
- LogTag® USRIC-8
- LogTag® USRIC-4
- LogTag® USRID-16
- LogTag® USRID-16W
- LogTag® UTRID-16
- LogTag® UHADO-16
- LogTag® UTRED30-16
- LogTag® UTRED30-WiFi
- LogTag® TRIL-8 / SRIL-8
- LogTag® TREL-8
- LogTag® TREL30-16
- LogTag® UTREL-16F
- LogTag® UTREL30-16
- LogTag® UTREL30-WiFi
- LogTag® TRID30-7
- LogTag® TRED30-16R
- LogTag® vaxtag
Sonde Externe LogTag
LogTag® TRED este proiectat si calibrat pentru a fi folosit cu sondele de temperatura LogTag® ST100Doar aceasta sonda este compatibila cu TRED si folosirea sa impreuna cu orice alta sonda va cauza erori in masurare.
Sondele LogTag® ST100 sint proiectate si calibrate pentru a fi interschimbabile si a pastra precizia de masurare. Daca o combinatie sonda TRED/ST100 este calibrata pentru un nivel inalt de precizie, atunci trebuie avut grija ca sa se foloseasca mereu aceeasi pereche TRED/ST100.
Identificarea unei sonde ST100
Sonda ST100 este distincta fata de alte tipuri de sonde externe LogTag, prin partea de culoare albastra din jurul conectorului.
Sonda scurta
ST100S-15 (1,5 m)
ST100S-30 (3,0 m)
Sonda scurta incorporeaza o lungime expusa de 22 mm, cu diametru de 5 mm din otel inox 316 fara miner, dar cu o mica adaugire din plastic, pentru a imbunatati rezistenta la umezeala si la indoire.
Toate materialele de la capatul sondei corespund cerintelor alimentare FDA, inclusiv cablul
Stainless probe Ø 3.2 mm, short
ST100J-15 (1.5 m)
ST100J-30 (3.0 m)
This short tipped probe incorporates ~40mm of exposed length of 3.2 mm diameter 316 stainless probe.
All materials at probe end of cable are FDA food rated including the cable.
Supplied standard (part# ST100J-15) with 1.5 m (4'11”) or extended (part#ST100J-30) with 3 m (9'10") cable terminated to MCX plug.
Sonda manevrabilă
ST100T-15 (1,5 m)
ST100T-30 (3,0 m)
Sonda simplă este dintr-un metal inoxidabil de 105 mm lungime si 5mm diametru, fără ,mâner însă cu o bună protecţie a contactului dintre cablu şi sondă. Toate materialele sondei şi cablul sunt aliniate standardelor impuse de FDA. Sonda ST100T-15 cu cablu de 1,5 m (4’11”) sau 3,0 m (9’10”) prevăzută cu conector MCX pentru conectarea la LogTag TREX.
Stainless probe Ø 3.2 mm, medium
ST100K-15 (1.5 m)
ST100K-30 (3.0 m)
This medium tipped probe incorporates ~65 mm of exposed length of 3.2 mm OD 316 stainless probe.
All materials at probe end of cable are FDA food rated including the cable.
Supplied standard (part# ST100K-15) with 1.5 m (4'11”) or extended (part#ST100K-30) with 3 m (9'10") cable terminated to MCX plug.
Sonda simplă
ST100H-15 (1,5 m)
ST100H-30 (3,0 m)
Sonda standard manevrabilă, este de 105 mm lungime si diametru de 5 mm din metal inoxidabil, prevăzută cu un mâner modelat de 70 mm. Toate materialele sondei şi cablul sunt aliniate standardelor impuse de FDA. Sonda ST100H-15 cu cablu de 1,5 m (4’11”) sau 3,0 m (9’10”) prevăzută cu conector MCX pentru conectarea la LogTag TREX.
Stainless probe Ø 3.2 mm, long
ST100L-15 (1.5 m)
ST100L-30 (3.0 m)
This long tipped probe incorporates ~90 mm of exposed length of 3.2 mm OD 316 stainless probe.
All materials at probe end of cable are FDA food rated including the cable.
Supplied standard (part# ST100L-15) with 1.5 m (4'11”) or extended (part#ST100L-30) with 3m (9'10") cable terminated to MCX plug.
Sondă neizolată
ST100B-15 (1,5 m)
ST100B-30 (3,0 m)
Senzorul fară înfăşurare metalică poate fi folosit oricând în sisteme dedicate. Nu este dedicat pentru contactul cu produse alimentare. Sonda ST100B-15 cu cablu de 1,5 m (4’11”) sau 3,0 m (9’10”) prevăzută cu conector MCX pentru conectarea la LogTag TREX
External sensor extension cable
STX-50 (5.0 m)
The STX-50 is a 5 meter (~16’ 5”) extension cable for extending ST10 or ST100 external temperature sensors.
The socket end of the cable (which accepts the ST100 or ST10 sensor plug) is fitted with a flexible sleeve that when the ST sensor’s plug is in place will protect the mated connection to IP61.
Data Sheet STX-50