LogTag® USRIC-8 / USRIC-16Plug and Play Temperature Monitoring

LogTag® USRIC-8 / USRIC-16Plug and Play Temperature Monitoring

Single-Use PDF USB Temperature Logger

The LogTag® USRIC-8 / USRIC-16 PDF Temperature Recorders measure and store up to 8,000 / 16,000 real time temperature readings over a measurement range of -30 °C to +60 °C

Applications

Data logger with integral USB connector. Ideal for cost-effective monitoring of temperature sensitive goods in both short haul and long-distance transportation applications

Product Summary

The LogTag® USRIC-8 PDF Temperature Recorder measures and stores up to 8,000 real time temperature readings over a measurement range of -30 °C to +60 °C.

The LogTag® USRIC-16 PDF Temperature Recorder measures and stores up to 16,000 real time temperature readings over a measurement range of -30 °C to +60 °C.

Designed as a single-use, cost effective & reliable temperature recorder, the LogTag® USRIC-8 / USRIC-16, enclosed in a robust & durable polycarbonate case, automatically generates a fully detailed PDF report and provides the option to download the data using the freely available LogTag Analyzer software.

Using LogTag® Analyzer, the LogTag® USRIC-8 / USRIC-16 is easily set-up for a number of recording conditions.

Product Highlights

  • Integral USB connector - No separate Interface Cradle required
  • Automatically generates a PDF report - No special software required
  • Encrypted LTD file generated for a more detailed analysis
  • User configurable - No need to buy a special logger for each trip
  • Suitable for independent receivers, clinical trials & last-mile distribution

Technical Specifications

Model USRIC-8 USRIC-16
Sensor Measurement Range -30 °C ~ +60 °C
Operating Temperature Range -30 °C ~ +60 °C
Storage Temperature Range 0 °C to +40 °C
Rated Temperature Reading Accuracy Better than ±0.5 °C for -5 °C to +30 °C
Better than ±0.8 °C or better for measurements in other areas of rated range
Actual performance is typically much better than the rated values.
Accuracy figures can be improved by recalibration.
Rated Temperature Reading Resolution < 0.1 °C
LogTag Analyzer currently displays to one decimal place of °C. The native resolution is what is stored in the LogTag®
Sensor Reaction Time Typically less than 5 minutes (T90) in moving air (1m/s) Typically less than 7 minutes (T90) in moving air (1m/s)
Recording Capacity 8,001 real time temperature recordings, non-volatile memory 16,129 real time temperature recordings, non-volatile memory
Sampling Interval Logging interval is configurable from 1 minute to 18 hours. (Typically 5 minute logging)
Logging Start Options Push button start or specific date & time. Optional start delay (1min to 72 hours)
Recording Indication Flashing 'OK' indicator / flashing 'ALERT' indicator.
Download Time Typically with full memory (8,000 readings) in less than 20 seconds from time of insertion to availability of PDF report.
Typically less than 10 seconds from time of insertion to availability of LTD file in LogTag® Analyzer (if configured)
Typically with full memory (16,129 readings) in less than 20 seconds from time of insertion to availability of PDF report.
Typically less than 10 seconds from time of insertion to availability of LTD file in LogTag® Analyzer (if configured)
Environmental IEC 60529: IP64 with USB cap fitted
Power Source 3 V Lithium battery– non replaceable
Download Time Typically in less than 20 seconds from time of insertion to availability of PDF report
Typically less than 10 seconds from time of insertion to availability of LTD file in LogTag® Analyzer (if configured)
Connection Interface USB 2.0, A-type plug
Software Adobe Reader, LogTag Analyzer® 2.5 or higher
Battery Life Projected life of 6 months of operation from time of configuration. Shelf life before configuration of up to 2 years.
Size 93 mm (H) x 54.5 mm (W) x 8.6mm (T) including protective USB cap.
Weight 32 g
Case Material Polycarbonate
LogTag® Recorders reserves the right to change this product’s specifications without notice.
EMC Compliance

Tested & complies with EC EMC directives (EN 50081-1:1992 & EN 61000-6-1:2001).
Tested & complies FCC Part 15 Subparts A & B.


FDA CFR21 Part 11 Compliance

Designed for FDA CFR21 Part 11 Digital signatures.


RoHS Directive

The RoHS Directive stands for "the restriction of the use of certain hazardous substances in electrical and electronic equipment". This Directive bans the placing on the EU market of new electrical and electronic equipment containing more than agreed levels of lead, cadmium, mercury, hexavalent chromium, polybrominated biphenyl (PBB) and polybrominated diphenyl ether (PBDE) flame retardants.