Kontakt

Questions? Price info? Technical details?

Iletişim Formu

LogTag® UTRID-16 Multi-Use PDF Logger at a glance

LogTag® UTRID-16 Multi-Use PDF Logger at a glance

Applications

Data logger with integral USB connector. The LogTag® UTRID-16 satisfies the growing need for a cost effective and reliable multi-use temperature recorder with built-in USB and display.

Product Summary

The LogTag UTRID-16 provides the facility to create a detailed PDF report without need for extra accessories and proprietary software and features a display, allowing immediate viewing of current and past temperatures as well as alarm statistics.

The display shows recording status, last recorded temperature and alarm status of up to six user-configurable alarms and duration of any temperature excursions. When the UTRID-16 is plugged into a computer's USB port using its integrated plug, a detailed PDF report of the recorded temperature data and alarms is be automatically generated.

For a more detailed data analysis, the UTRID-16 also creates an encrypted data file compatible with LogTag Analyzer® software and data management options.

Product Highlights

  • Display showing temperature & alarm status of up to six alarms.
  • Can be connected to any computer with a USB port.
  • Automatic generation of a PDF report as a permanent record of the data.
  • Creates an encrypted data file compatible with LogTag Analyzer.
  • UTRID-16 offers both a user Replaceable battery and a Fixed Battery type.
  • Ideal for multi-stage distribution channels involving “last-mile”.
  • UTRID-16 is Pre-Qualified compliant to WHO PQS specification E006 Temperature monitoring devices

Technical Specifications

Model UTRID-16R (Replaceable Battery).
UTRID-16F (Fixed Battery) WHO PQS E006/076 qualified
Measurement Range -30 °C to +60 °C
Capacity 16,129 temperature readings.
112 days @ 10 min logging, 168 days @ 15 min logging. Statistics memory (for display on LCD): Max/Min values and alarm duration values.
Sampling Interval Configurable from 30 seconds to 18 hours.
Start Options Push button start or specific date/time.
Start Up Delay Optional start delay of up to 72 hours.
Environmental IEC 60529: IP 64 with USB cap fitted.
Power Source UTRID-16R - CR2032 3V LiMnO2 Battery (Replaceable).
UTRID-16F - CR2450 3V LiMnO2 UN38.3 approved coin cell battery (Fixed)
Battery Life

UTRID-16R: 1 year storage life before ‘start’ (if it is hibernation mode), followed by: 1 year of normal use (based on 15 minute logging, download data monthly).
UTRID-16F: Typically 2 years of operation with normal use (6 minute logging, statistics reviewed on the display no more than once daily for no longer than 30 seconds each time, download data monthly) provided recorder is kept within the storage temperature range when not in use.

Size 93 mm(H) x 54.5 mm (W) x 8.6 mm (T) including protective USB cap.
Weight 39 g
Case Material Polycarbonate
Alarm Functions Three configurable upper and three configurable lower alarms.
Connection Interface USB 2.0, A-type plug.
Software PDF Reader, LogTag® Analyzer 2.8 or higher.
LogTag® Recorders reserves the right to change this product’s specifications without notice.
EMC Compliance

Tested & complies with EC EMC directives (EN 50081-1:1992 & EN 61000-6-1:2001).
Tested & complies FCC Part 15 Subparts A & B.


FDA CFR21 Part 11 Compliance

Designed for FDA CFR21 Part 11 Digital signatures.


RoHS Directive

The RoHS Directive stands for "the restriction of the use of certain hazardous substances in electrical and electronic equipment". This Directive bans the placing on the EU market of new electrical and electronic equipment containing more than agreed levels of lead, cadmium, mercury, hexavalent chromium, polybrominated biphenyl (PBB) and polybrominated diphenyl ether (PBDE) flame retardants.