Highlights

  • Pre-Qualified compliant to WHO PQS specification E006/070 Temperature monitoring devices
  • Stores 40 days of temperature recordings data, factory set with a 5 minute logging interval.
  • Display showing temperature and alarm status of up to three alarms.
  • Automatic generation of a PDF, CSV, LTD report without special software.

Data logger with integral USB connector. The LogTag® WHO compliant USRID-16WC and USRID-16WA/B satisfy the growing need for a cost effective and reliable single-use temperature recorder with built-in USB and display. They are ideally suited for vaccine transportation, pharmaceutical transportation as well as in laboratories and medical centres & healthcare.

The display shows alarm status of up to 3 alarms set at WHO requirements. It plugs directly into the computer USB for easy data download.
Upon download, it automatically generates a fully detailed PDF report with the option to download the data using the freely available LogTag® Analyzer software for a more detailed analysis.

 

USRID-16WC
USRID-16W1

WHO Type 1 alarm configuration for freeze sensitive vaccines:
DTP, DT
TT, Td
HepB, IPV, HPV
Liquid Hib and combination vaccines
Pneumo (other than Prevenar)

USRID-16WA/B
USRID-16W2

WHO Type 2 alarm configuration suitable for the international shipment of:
BCG, lyophilized Hib, OPV
Measles, MR, MMR, Rabies,
Meningitis (polysaccharide)
Yellow fever vaccines
Rotavirus (other than rotateq)

Product Model USRID-16W
Temperature Sensor Measurement range -30 °C to +60 °C
Recorder Operating Temperature Range -30 °C to +60 °C
Storage Temperature Range -30 °C to +55 °C with device de-activated
Rated Temperature Reading Accuracy Better than ±0.5 °C for -5 °C to +30 °C
Better than ±0.8 °C or better for -30 °C to +60 °C
Actual performance is typically much better than the rated values. Accuracy figures can be improved by recalibration
Rated Temperature Reading Resolution Better than 0.1 °C between -20 °C to +55 °C
LogTag Analyzer currently displays to one decimal place of °C.
The native resolution is what is stored in the LogTag
Sensor Reaction Time Typically less than 7 minutes (T90) in moving air (1m/s)
Recording Capacity 11,520 temperature readings (40 days at 5 minute logging)
Sampling Interval Factory set at 5 minutes (non-changeable)
Logging Start Options Push button start
Alarms 3 alarms set at WHO requirements
Download Time Typically with full memory in less than 20 seconds from time of insertion to availability of PDF report
Typically less than 10 seconds from time of insertion to availability of LTD file in LogTag® Analyzer 3
Environmental Ingress rating: IP65 with USB cap fitted Shock and Vibration: Unit shall pass a relevant level of shock and vibration based on the requirements of the Indicator specification E006/ TR07.3 Environmental resistance: Labels and plastics shall be resistant to a range of chemicals and UV, including rubbing alcohol and petrol fuel
Power source CR2032 3V LiMnO2 Battery (Non-Replaceable)
Battery Life Minimum storage life of 18 months before ‘start’
Monitoring period: 40 days Minimum accessibility (display) period of 6 months after ‘stopped’
Real Time Clock Built-in real time clock Rated accuracy ±25ppm @ 25°C (equivalent to 2.5 seconds/day)
Rated temperature coefficient is -0.034 ±0.006ppm/°C (I.e typically +/- 0.00294 seconds/day/°C)
Connection Interface USB 2.0, A-type plug
Software PDF Reader, LogTag Analyzer 3 or higher
Size 93 mm(H) x 54.5 mm(W) x 8.6 mm(T) including protective USB cap
Weight 39g
Case Material Polycarbonate
Logging Stop Options Push button stop Stop after the logger has completed 11,520 temperature readings (40 days at 5 minute logging)