The device may also be used as a secondary back-up device in cold rooms. The product features a display together with a data logging function storing up to 7770 temperature readings. Statistical temperature and duration readings for up to 30 days can be reviewed on the display.
Details of any violations can be checked directly by inspecting the statistics history on the recorder’s display or in more detail by downloading the logged data.Logged data can be downloaded via a standard LogTag® Interface to the free companion software LogTag Analyzer which provides facilities for displaying data in chart, table or statistical formats and allows electronic archiving, export or transmission of the data in support of sophisticated data management systems.
The device is preconfigured with the following alarm settings:
• Low alarm: Exposure to -0.5 °C or below for 60 minutes.
• High alarm: Exposure to +8 °C or above for 10 hours.
A visual alarm trigger is displayed if one or more of the configured alarm trigger conditions have been met.
Details of any violations can be checked directly by inspecting the statistics history on the recorder’s display or in more detail by downloading the logged.
Please note: An interface is required for configuration and reporting
Product Model |
TRID30-7FW (Vaxtag) |
Temperature Sensor Measurement range |
-30 °C to +60 °C (-22 °F to +140 °F) |
Recorder Operating Temperature Range |
-30 °C to +60 °C (-22 °F to +140 °F) |
Storage Temperature Range |
-10 °C to +40 °C (-14 °F to +104 °F) |
Rated Temperature Reading Accuracy |
Better than ±0.5 °C (±0.9 °F) for measurements from -20 °C to +40 °C (-4 °F to +104 °F), typically ±0.3 °C (0.6 °F). Better than ±0.8 °C (±1.5 °F) for other measurements - typically ±0.5 °C (0.9 °F). Actual performance is typically much better than the rated values. Accuracy figures can be improved by recalibration |
Rated Temperature Reading Resolution |
0.1 °C(0.2 °F) for measurements -30 °C to +40 °C (-22 °F to +104 °F). 0.2 °C(0.4 °F) for measurements above +40 °C (+104 °F). LogTag Analyzer® currently displays to one decimal place of °C or °F. The native resolution is what is stored in the LogTag |
Sensor Reaction Time |
Typically less than 5 minutes (T90) in moving air (1m/s) |
Recording Capacity |
7770 temperature readings. 32 days @ 6min logging Day summary statistics memory (for display on LCD): up for 30 days of Max/Min and duration values |
Sampling Interval |
Factory set to 6 minute interval |
Logging Start Options |
Factory set to Push button start |
Recording Indication |
State indicator “RECORDING” |
Alarms |
Low alarm : Exposure to a single temperature event of -0.5°C or below for 60 minutes High alarm: Exposure to a single temperature event of +8°C or above for 10 hours |
Download Time |
Typically less than 5 seconds for full memory (7770 readings), depending on computer or readout device used |
Environmental |
IP65 (roughly equivalent to NEMA 4) when vertically mounted or hung |
Power source |
CR2450 3V LiMnO2 (Fixed Battery) |
Battery Life |
1 year storage, followed by 2 year of operation with normal use (6 minute logging, statistics reviewed on the display no more than once daily for no longer than 30 seconds each time, download data monthly) |
Real Time Clock |
Built-in real time clock. Rated accuracy ±25ppm @ 25°C (equivalent to 2.5 seconds/day). Rated temperature coefficient is -0.034 ±0.006ppm/°C (I.e typically +/- 0.00294 seconds/day/°C) |
Size |
93 mm(H) x 54.5 mm(W) x 8.6 mm(T) |
Weight |
43 g |
Case Material |
Polycarbonate |
Wall Mount Bracket WMB-01